An independent comparison study was conducted by the WSLH Proficiency
Testing Program. HemaPrompt FG and Gastroccult (20 slides of each
product) were used to compare sensitivity to Hemoglobin in synthetic
gastric juice (manufactured by WSLH Proficiency Testing ) over a
range of Hgb concentrations (500ug/ml minimum) and pH range 1 -
7.
No significant differences were found between the two products and
each was found equally suitable for the detection of Hemoglobin
in gastric juice in the pH range of 1 to 7. ( September 2003)
|
|